Description Code
- 125 ml anti-spill vessel ACO48
- Ø 4 x 6 mm connection tubes, 160 mm long
- Ø 4 x 6 mm connection tube, 1000 mm long
- 250 ml collection vessel
- Disposable CH 18 Ø 6 x 4.4 mm sterile cannula, 530 mm long 06488
Use only original Flaem accessories
AVAILABLE SPARE PARTS
The following accessories can be purchased from your dealer or an authorised after-sales centre.
u
Set of 2 antibacterial lters - Code ACO376P
OPTIONAL ACCESSORIES (to be purchased separately)
Filter operating instructions
By means of the connection tube provided with the lter connect the air
inlet port (A3) of the unit to an end of the lter and the short tube (B1)
provided with the unit to the other end of the lter (see gure 1).
- Remove the unit from the carrying bag before use.
- The antibacterial lter is disposable.
u
Carrying bag - Code 11438
Filter/ air inlet
connection tube
Filter
SANITIZATION
Before and after each use, sanitise the accessories by choosing one of the methods described below
(method A): Sanitize the accessories (C1-C2-C3) using warm tap water (40°C approx.) with mild (nonabrasive) washing-up liquid or
wash them in a dishwasher, running a hot cycle.
(method B): Sanitize the accessories (C1-C2-C3) by soaking them in a solution of 50% water and 50% white vinegar. Rinse thoroughly
with warm tap water (40°C approx).
Drying: After sanitizing the accessories, shake them vigorously and lay them on a paper towel or dry them with a jet of warm air (eg
a hair dryer).
DISINFECTION
(method A): The accessories that can be disinfected with method A are (C1-C2-C3)
The disinfection procedure described in this paragraph is to be performed before using the accessories and is effective only if all
the prescriptions herein are met and all the components involved are previously sanitized. The disinfectant to be used must be of
electrolytic oxidizing chlorine type, (active ingredient: sodium hypochlorite), specic for disinfection and is available in all pharmacies.
Procedure:
- Fill a container, suitable to contain all the individual components to be disinfected, with a solution of drinking water and disinfectant,
while respecting the proportions indicated on the packaging of the disinfectant itself.
- Completely immerse each individual component in the solution, taking care to avoid the formation of air bubbles in contact with the
components. Leave the components immersed for the period of time indicated on the packaging of the disinfectant, and associated
with the concentration chosen to prepare the solution.
- Remove the components now disinfected and rinse thoroughly with lukewarm drinking water.
- After sanitization shake the accessories vigorously and place them on a paper towel, or alternatively, dry them with a jet of hot air
(e.g. using a hair dryer).
- Dispose of the solution as recommended by the disinfectant manufacturer.
(method B): The accessories that can be disinfected with method B are (C1-C2-C3)
Disinfecte the accessories by boiling them in water for 20 minutes; use demineralised or distilled water to prevent limescale deposits.
- After disinfection shake the accessories vigorously and place them on a paper towel, or alternatively, dry them with a jet of hot air
(e.g. using a hair dryer).
FIG. 1