Operation
11
11.3.1 Automatic temperature range selection
11.3.1.1 General
The automatic temperature range selection is based on a measured area defined by a rec-
tangle covering (x1, y1) = (15% of the width, 15% of the height) to (x2, y2) = (85% of the
width, 85% of the height) of the LCD area. See the figure in section 11.3.1.2.
An automatic change from the high-sensitivity range to the low-sensitivity range occurs if
more than 2% of the pixels within the measured area constantly (for more than 1 second)
have a temperature above the maximum temperature of the high-sensitivity range.
An automatic change from low-sensitivity range to high-sensitivity range occurs if more
than 98% of the pixels within the measured area constantly have, for more than 1 second,
a temperature lower than 50°C (122°F) below the maximum temperature of the high-sensi-
tivity range.
11.3.1.2 Figure
11.3.1.3 Explanation
1. LCD area.
2. Area activating the automatic range change.
11.4 Saving an image (FLIR K53)
11.4.1 General
You can save images to the camera’s archive.
Note The maximum number of images that can be saved in the archive is 200. When
the number of images exceeds 200, images are deleted on a first-in, first-out basis, i.e.,
the 201st image will delete the 1st image, the 202nd image will delete the 2nd image, and
so on.