ThermaCAM™ SC640 45° Package
P/N: 30901-0303
© 2017, FLIR Systems, Inc.
#30901-0303; r. /29700; en-US
Image presentation modes
Infrared image Full IR-image with selected color scale
Visual image Full color visual image
Reference image Shown together with live IR image
Measurement
Object temperature range –40°C to +120°C (–40°F to +248°F)
0°C to +500°C (+32°F to +932°F)
+350°C to +1500°C (+662°F to +2732°F)
Accuracy ±2°C (±3.6°F) or ±2% of reading
Measurement analysis
Spotmeter
10
Area 5 boxes or circles with max./min./average
Profile 1 live line (horizontal or vertical)
Automatic hot/cold detection Max/Min temp. value and position shown within
box, circle or on a line
Isotherm
2 with above/below/interval
Difference temperature Delta temperature between measurement
functions or reference temperature
Reference temperature Manually set or captured from any measurement
function
Atmospheric transmission correction Automatic, based on inputs for distance,
atmospheric temperature and relative humidity
Optics transmission correction Automatic, based on signals from internal
sensors
Emissivity correction Variable from 0.01 to 1.0 or selected from
editable materials list
Emissivity table Emissivity table of predefined and editable
materials
Reflected apparent temperature correction Automatic, based on input of reflected
temperature
External optics/windows correction Automatic, based on inputs of optics/window
transmission and temperature
Alarm
Measurement function alarm Audible/visual alarms (above/below) on any
selected measurement function
Set-up
Set-up commands Configurable measurment tools menu; configure
information to be shown in image; 2
Programmable buttons; user profiles; local
adaptation of units, language, date and time
formats
Storage of images
Storage media Removable memory card
Built-in RAM memory for burst recording
Image storage mode
IR/visual images, simultaneous storage of IR and
visual images.
Visual image is automatically associated with
corresponding IR image.
Periodic image storage Every 10 seconds up to 24 hours