Guidance and Manufacturer’s Declaration – Electromagnetic
Emissions – for all EQUIPMENT and SYSTEMS
Guidance and Manufacturer’s Declaration – Electromagnetic
Immunity – for all EQUIPMENT and SYSTEMS
Guidance and Manufacturer's Declaration - Electromagnetic Emissions
The Infrared Thermometer is intended for use in the electromagnetic
environment specified below. The customer or user of the device should
ensure that it is only used in such and environment.
Emissions
Test
Compliance
Electromagnetic Environment - Guidance
As this device only uses RF energy for its
internal functions, its RF emissions are very
low and not likely to cause any interference to
nearby electronic equipement.
This device is suitable for use in all establish-
ments, including domestic establishments and
those directly connected to the public
low-voltage power supply network which
supplies buildings for domestic purposes.
Group 1
Class B
RF Emissions
CISPR 11
RF Emissions
CISPR 11
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The Infrared Thermometer is intended for use in the electromagnetic
environment specified below. The customer or user of the device should
ensure that it is only used in such and environment.
Immunity
Test
IEC 60601
Test Level
Electromagnetic Environment -
Guidance
Floors should be wooden,
concrete, or ceramic tile. If floors
are covered with synthetic
material, the relative humidity
should be at least 30%
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environement
Compliance
Level
+/- 6KV
contact,
+/- 8KV air
+/- 6KV
contact,
+/- 8KV air
3 A/m 3 A/m
Electrostatic
Discharge
(ESD) IEC
61000-4-2
Power
Frequency
(50/60 Hz)
Magnetic Field
IEC 61000-4-8
- 19 -