EasyManua.ls Logo

Fluke 77 - Page 18

Fluke 77
20 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Model 77 Series IV
Users Manual
14
Function Overload Protection
[1]
Input Impedance
(Nominal)
Common Mode Rejection Ratio
(1 k Unbalanced)
Normal Mode Rejection
Volts AC 1000 V >10 M <100 pF >60 dB @ dc, 50 Hz or 60 Hz
Volts DC 1000 V >10 M <100 pF >120 dB @ dc, 50 Hz or 60 Hz >60 dB @ 50 Hz or 60 Hz
mV 1000 V
[2]
>10 M <100 pF >120 dB @ dc, 50 Hz or 60 Hz >60 dB @ 50 Hz or 60 Hz
Open Circuit Test
Voltage
Full Scale Voltage To:
6.0 M 50 M
Short Circuit Current
Ohms/Capacitance 1000 V
[2]
<8.0 V dc <660 mV dc <4.6 V dc <1.1 mA
Continuity/Diode test 1000 V
[2]
<8.0 V dc 2.4 V dc <1.1 mA
[1] 10
7
V-Hz maximum.
[2] For circuits <0.3 A short circuit. 660 V for high energy circuits.
Function Overload Protection Overload
mA Fused, 440 mA, 1000 V FAST Fuse 600 mA overload for 2 minutes maximum, 10
minutes rest.
A Fused, 11 A, 1000 V FAST Fuse 20 A overload for 30 seconds maximum, 10
minutes rest.

Other manuals for Fluke 77

Related product manuals