Specifications
The 80T-150UA will achieve rated accuracy when it is used with any
0.25 % DMM that has an input impedance of 1 MΩ.
Electrical
Measurement Range: (See Probe Limitations)
-50 to +150 °C
-58 to +302 °F
Accuracy: (Applies for one year after purchase or calibration)
AMBIENT °C ACCURACY
+15 to +35 °C ±1 °C from 0 to +100 °C, decreasing
linearly to ±3 °C at -50 and +150 °C
0 to +15 °C and ±2 °C from 0 to +100 °C,
+35 to +50 °C decreasing linearly to ±4 °C at -50
and +150 °C
AMBIENT °F ACCURACY
+59 to +95 °F ±1.8 °F from +32 to +212 °F,
decreasing linearly to ±5.4 °F at -58
and +302 °F
+32 to +59 °F and ±3.6 °F from +32 to +212 °F,
+95 to +122 °F decreasing linearly to ±7.2 °F at -58
and +302 °F
Sensitivity (80T-150UA output): 1 mV dc / °C or °F
Voltage Standoff: 60 V dc or 30 V ac rms (42.4 V peak)
Settling Time: 9.5 seconds to settle within 2 ° for a 50 ° change
RF Field: The 80T-150UA product is unspecified in environments with
EMC fields greater than 100 mV/meter. ESD performance is specified as
Criteria C per EN61326.
Environmental
Ambient Operating Range for Temperature Module: 0 to +50 °C
(+32 to +122 °F)
Maximum Temperature Probe Body and Cable: +70 °C
(+160 °F) See Probe Limitations
Minimum Temperature Probe Body and Cable:
Minimum Operating Temperature for Cable: -12 °C (+10 °F)
Minimum Operating Temperature for Probe Body: -40 °C (-40 °F)
See Probe Limitations
Storage Temperature for Unit: -40 to +70 °C (-40 to +160 °F)
Humidity: 0 % to 90 % (0 °C to +35 °C)
0 % to 70 % (+35 °C to +50 °C) (noncondensing)
Altitude: Operating: 3048 m ( 10,000 ft)
Storage: 15240 m (50,000 ft)
Application Force: 15 pounds maximum (probe tip to measured surface)