5 - 18 Users Manual
Remote control (in CV or CC mode):
The volta
g
e V
set
can be pro
g
rammed by the SOURCE subsystem command:
[:SOURce]:VOLTa
g
e[:LEVel][:IMMediate][:AMPLitude]
The volta
g
e V
set
can be requested by the SOURCE subsystem query:
[:SOURce]:VOLTa
g
e[:LEVel][:IMMediate][:AMPLitude]?
The output volta
g
e can be read back by the MEASURE subsystem query:
:MEASure[:SCALar]:VOLTa
g
e[:DC]?
Example:
Send :VOLT 3.4 Set 3.4V.
Send :VOLT? Send request volta
g
e settin
g
.
Read set_volta
g
e Read volta
g
e settin
g
.
Print set_volta
g
e Display volta
g
e settin
g
.
Send :MEAS:VOLT? Send readback query.
Read meas_volta
g
e Read output volta
g
e.
Print meas_volta
g
e Display measured volta
g
e.
The current I
set
can be pro
g
rammed by the SOURCE subsystem command:
[:SOURce]:CURRent[:LEVel][:IMMediate][:AMPLitude]
The current I
set
can be requested by the SOURCE subsystem query:
[:SOURce]:CURRent[:LEVel][:IMMediate][:AMPLitude]?
The output current can be read back by the MEASURE subsystem query:
:MEASure[:SCALar]:CURRent[:DC]?
Example:
Send :CURR 0.23 Set 0.23A.
Send :CURR? Send request current settin
g
.
Read set_current Read current settin
g
.
Print set_current Display current settin
g
.
Send :MEAS:CURR? Send readback query.
Read meas_current Read output current.
Print meas_current Display measured current.