Calibration Manual
18
Disassembly
XWWarning
To prevent possible electrical shock or personal injury,
disconnect the mains power cord before you remove the
Product covers.
Only disassemble the Product to reset the password or replace the battery. See
Figure 8.
A 2# Phillips screwdriver and small crescent wrench are required for
disassembly.
1. Remove all High-Capacity Modules and test leads from the Product.
2. Turn off the mains power at the rear of the Product and remove the mains
power cord. The front panel power key only puts the Product into a power-
save mode and does not remove mains power.
3. Remove the bail by pulling from a corner and stretching the boot off the
Product.
4. Remove the bail by rotating the handle upright to a 90 ° angle from the top of
the Product and pull the bail out from the sides.
5. Remove the top cover by removing the two screws on the sides of the
chassis and slide the cover towards the back of the Product.