1586A
Calibration Manual
2
General Maintenance
The subsequent sections describe how to maintain the Product.
Clean the Product
To clean the Product, wipe the instrument with a cloth that is lightly dampened
with water or mild detergent. Do not use aromatic hydrocarbons, chlorinated
solvents, or methanol-based fluids.
WCaution
To prevent possible damage to the Product, do not use
solvents or abrasive cleansers.
WCaution
For safe operation and maintenance of the Product:
• Have an approved technician repair the Product.
• Do not allow water to get inside the Product.
If the Product Does Not Turn On
To help solve problems encountered when turning on the Product:
1. Make sure that the power switch is in the “On” position.
2. Make sure that the mains power cord is firmly plugged into the power module
on the rear of the Product.
3. Make sure the power source that the Product is plugged into is energized.
4. Make sure that the line power fuse is good. See Replace the Line Power
Fuse.
If these steps do not solve the problem, then contact Fluke Calibration. See
Contact Fluke Calibration.
Replace the Line Power Fuse
To replace the fuse, see the Users Manual.