Test
frequency
(MHz)
385
450
2 450
710
745
780
5 240
5 500
5 785
810
870
930
1 720
1 845
1 970
380 –390
430 – 470
704 – 787
800 – 960
1 700 –
1 990
2 400 –
2 570
5 100 –
5 800
TETRA 400
GMRS 460,
FRS 460
LTE
Band 13,
17
GSM
800/900,
TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5
GSM 1800;
CDMA 1900;
GSM 1900;
DECT;
LTE
Band 1, 3,
4, 25; UMTS
Bluetooth,
WLAN,
802.11 b/g/n,
RFID 2450,
LTE Band 7
WLAN 802.11
a/n
Pulse
modulation b)
18 Hz
Pulse
modulation b)
217 Hz
Pulse
modulation b)
217 Hz
Pulse
modulation b)
217 Hz
Pulse
modulation b)
18 Hz
FM c)
±5 kHz
deviation
1 kHz sine
Pulse
modulation b)
217 Hz
1,8
2
2
0,2
0,3 27 27
0,3 28 28
0,3 28 28
2
0,3
0,3
28 28
2
0,30,2
28 28
0,3 9 9
9 9
Band a)
(MHz)
Service a) Modulation b) Maximum power
(W)
Distance
(m)
IMMUNITY TEST
LEVEL
(V/m)
Compliance LEVEL
(V/m)
(for home healthcare)
Manufacturer’s declara�on-electromagne�c immunity
Test specifica�ons for ENCLOSURE PORT IMMUNITY to RF wireless communica�ons equipment
The IR42 is intended for use in the electromagnetic environment (for home healthcare)
specified below.
The customer or the user of the IR42 should assure that it is used in such an environment.
NOTE If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting antenna and the ME
EQUIPMENT or ME SYSTEM may be reduced to 1 m. The 1 m test distance is permitted by IEC 61000-4-3.
a) For some services, only the uplink frequencies are included.
b) The carrier shall be modulated using a 50 % duty cycle square wave signal.
c) As an alternative to FM modulation, 50 % pulse modulation at 18 Hz may be used because while it does not
represent actual modulation, it would be worst case.