Model No.: IR42
Dimension & Weight: 155.46 (L) x 40.14 (W) x 39.45 (H) mm,
61.8g (without battery)
Power Source: 2 x 1.5V AAA alkaline batteries
Battery Life: With new batteries, approx. 3,000 measurements.
Displayed Temperature range: 89.6°F to 109.4°F (32°C to 43°C)
Display Resolution: 0.1°F/0.1°C
Accuracy: Meet the accuracy requirement specified in ASTM
E1965-98
±0.4°F (±0.2°C) for the range of 95°F to 107.6°F (35.0°C to 42.0°C)
/ ±0.5°F (±0.3°C) for the range of <95°F (35.0°C) or >107.6°F
(42.0°C)
Reference to Standards: ASTM E1965-98; IEC 60601-1; IEC
60601-1-2 (EMC)
Temperature unit: °F (Default) or °C
Operating temperature range: 50°F to 104°F (10°C to 40°C)
Operating humidity: 85% RH or less
Storage / transportation temperature range: -4°F to 140°F (-20°C
to 60°C)
Storage / transportation humidity: 85% RH or less
Memory capacity: 30 measurements
Expected service life: 3 years
The specifications may be changed without prior notice.
TROUBLESHOOTING
What it meansMessage What to do
Appear when
ambient temperature
is outside of the
operating
temperature range.
Only operate the
thermometer in
ambient
temperature
range of 50°F to
104°F (10°C to
40°C).
Measured temperature
below the measurement
range (below 89.6°F).
Review instructions and
repeat measurement.
Make sure distance
from forehead is not
greater than 2”, and
that the forehead is
clear.
Measured temperature
above the measurement
range (above 109.4°F).
Surface temperature
measured outside of the
measurement range
(32.0°F ~ 212.2°F).
Review the instructions
and re-start the
measurement
procedure.
Low or no power. Replace with new
batteries.
Review instructions and
repeat measurement.
Consult with a physician
if the problem persists.
Distributed by ForaCare, Inc.
893 Patriot Drive Suite D
Moorpark, CA 93021 USA
Products made in Taiwan
Toll Free: 1-888-307-8188
(8:30 am - 5:00 pm PST, Mon.-Fri.)
For assistance outside of these hours,
please contact your healthcare professional.
www.foracare.com
Read instructions before use.