Calibration
Enter the Program settings, use SMPL key to chose the
item (CAL-Fun-04), then press CHECK SET key to enter,
use SMPL key to make the number to 1, then press
CHECK SET key, the LCD display inPut, use Numeric
keys(0-9) to set the calibration weight.
For example, 1000g mass weight, you need make the
number become 1000, press CHECK SET key, when the
number stable, the LCD will display the calibration weight
1000, then put 1000g mass weight on the pan, when the
number become stable, press CHECK SET key. Then
calibration is finished.
Recharge
1. After turning on the scale or in usage, the figures keep
shinning, which means the power of battery is not full. It