OPERATOR’S GUIDE
Volumetric
Infusion Pump
OPTIMA
PT
Introduction
Great care has been taken in the design and manufacture of the
Optima PT,
a volumetric infusion device.
The configuration flexibility of the Optima PT provides overall
improvement in the working conditions of medical teams, thus increasing
patient safety.
The use of this material requires great care. The user must be able to
handle the instrument properly and must know how to fully operate.
Please read the operator’s guide carefully before putting the device into
use.
Table of contents
1. Operations for use ........................................................................ 2
2. Advanced functions ...................................................................... 4
3. Internal safety features ................................................................. 7
4. Performances .............................................................................. 10
5. Optima PT - Technical characteristics ...................................... 11
6. Configuration menu .................................................................... 13
7. Intended Use................................................................................ 15
8. Precautions to be taken .............................................................. 15
9. Guidance and manufacturer's declaration on electromagnetic
environment ........................................................................................... 16
10. Maintenance recommendations ................................................ 18
11. RS 232 connection ...................................................................... 19
12. Nurse call connection ................................................................. 19
13. Function with the internal battery ............................................. 19
14. Accessories ................................................................................. 20
15. Disposable ................................................................................... 21
16. Conditions of guarantee ............................................................. 22
17. Useful addresses ........................................................................ 24
Question and Answer IconNeed help?

Do you have a question about the Fresenius Kabi OPTIMA PT and is the answer not in the manual?

Overview

The Optima PT is a volumetric infusion pump designed for medical use, particularly in settings where precise and controlled fluid administration is critical. It is manufactured by Fresenius Kabi and is compliant with Medical Device Directive 93/42/EEC.

Function Description

The Optima PT is an infusion pump that delivers fluids to patients at a controlled rate and volume. It supports various infusion modes, including continuous infusion, secondary infusion, and bolus administration. The pump is designed for use with specific infusion sets, ensuring accurate and safe delivery. It features a user-friendly interface with a display that shows critical information such as infusion rate, total volume infused, and remaining volume. The pump incorporates safety features like air-in-line detection, occlusion detection, and pre-alarm functions to alert users to potential issues during infusion. It can operate on both mains power and an internal battery, providing flexibility for different clinical environments.

Important Technical Specifications

  • Mains Supply: 100-240 V ~ 50/60 Hz
  • Max. Power Consumption: 50 VA
  • Internal Protective Fuse: 630 mAT
  • Battery: 6 V = 2.7/3 Ah - NiMH. Battery life is approximately 4 hours for any flow rate lower than 125 ml/h and 25 ml/h. Battery recharging takes approximately 5 hours.
  • Dimensions (Height/Width/Depth): 13.5 x 17.5 x 14.5 cm
  • Weight: Approximately 2.9 kg
  • Flow Rate Range: 0.1 ml/h to 1200 ml/h, with increments of 0.1 ml/h from 0.1 to 100 ml/h, and 1 ml/h from 100 to 1200 ml/h.
  • Volume Infused Range: 0.1 ml to 9999 ml, with increments of 0.1 ml.
  • Time Range: 0h01' to 96h00', with increments of 1 minute.
  • Pressure Limit: Adjustable from 100 to 900 mmHg.
  • Air Detection: Adjustable from 250 µl to 1000 µl, with a default setting of 50 µl.
  • Occlusion Detection: Pressure threshold for downstream occlusion is adjustable.
  • RS 232 Connection: Available for data communication.
  • Nurse Call Connection: Available for integration with hospital nurse call systems.
  • Protection Against Ingress of Water: IP31 (protection against dripping water).
  • Electrical Safety: Class I, Type CF equipment (defibrillation-proof applied part).

Usage Features

The Optima PT offers a comprehensive set of features designed to simplify and secure the infusion process:

  • Tubing Set Installation: A clear, step-by-step process for installing the infusion set, including priming the set and inserting it into the air sensor.
  • Setting Infusion Parameters: Users can easily set the volume to infuse, time, and flow rate. The pump automatically calculates the third parameter based on the other two. Drug names can also be programmed.
  • Infusion Modes: Supports various modes including continuous infusion, secondary infusion for administering additional solutions, and bolus administration for rapid drug delivery.
  • Programming Pause: Allows users to temporarily halt an infusion for a specified duration.
  • Key Locked/Unlocked: A feature to prevent accidental changes to infusion parameters during operation.
  • Loading Dose: Facilitates the delivery of a volume (e.g., loading dose) automatically followed by the delivery of the primary infusion.
  • Ramp Up/Ramp Down: Enables gradual increase and decrease of flow rate at the beginning and end of an infusion, respectively.
  • Sequential Programming: Allows programming up to 20 infusion sequences, each with different parameters, for complex infusion protocols.
  • Advanced Functions: Includes pressure monitoring, battery charge level display, and various alarm settings.
  • Display: An LCD screen provides clear visibility of all relevant infusion data and alarm messages.
  • Accessories: Compatible with accessories such as infusion pump rolling stands, bag/bottle holders, and drop sensors to enhance usability and flexibility in different clinical settings.

Maintenance Features

Regular maintenance is crucial for the Optima PT to ensure its continued accuracy and safety.

  • Cleaning and Disinfecting: The pump's external surfaces should be cleaned and disinfected regularly using recommended cleaning agents. It is important to disconnect the device from the power supply before cleaning.
  • Preventive Maintenance: Fresenius Kabi recommends preventive maintenance every 3 years. This includes battery replacement and should be performed by qualified technicians.
  • Quick Check: A daily or pre-use quick check procedure is outlined, covering visual inspection, LED and buzzer tests, pump door operation, tubing insertion, air sensor function, and occlusion alarm tests. This helps ensure the device is functioning correctly before each use.
  • Battery Maintenance: The battery life can decrease over time. For frequent mains operations, battery life may be extended by fully discharging and recharging the battery approximately every 4 weeks. A pre-alarm battery signal indicates when the battery needs attention.
  • Technical Service: For any technical issues, repairs, or comprehensive maintenance, users are advised to contact Fresenius Kabi's Technical Service. The device should be packed in its original packaging when sent for service.
  • Disposal: Batteries and devices must be disposed of according to local regulations, as they contain materials that could be harmful to the environment. Fresenius Kabi provides information on proper disposal.