897N200877B
5-3
Chapter 5 How to Use
5.1 Preparation
5.1.1 Preparing Related Equipment
Prepare related equipment to be used, including back-up or spare endoscope, scope
components (valves, distal caps, etc.) and accessories (ex. water tank, endotherapy devices,
etc.)
5.1.2 Pretreatment of Patient
Prepare the patient in the normal endoscopy regimen.
5.1.3 Preparing Mouthpiece
Prepare a mouthpiece per manufacturer’s instructions.
WARNING
• Use an appropriate mouthpiece as per manufacturer's instructions, including specific
reprocessing recommendations.
CAUTION
• Do not use a mouthpiece that is damaged, deformed, or reveals other irregularities. Doing
so may cause injury in the oral cavity and/or equipment failure.
Note Reusable mouthpieces are considered consumable items. If any abnormality is found,
use a new mouthpiece per manufacturer's instructions.
(1) Before oral insertion, have the patient hold the mouthpiece in his/her mouth.
Note If you choose to have the patient hold the mouthpiece after insertion, attach the
mouthpiece to the insertion portion in advance. Have the patient hold it promptly after
insertion.