Chapter 5 How to Use
5.2.3 Operating Scope Switch
Scope Switch “3”
Scope Switch “4”
Scope Switch “2”
Scope Switch “1”
Note Use the processor to assign functions on each scope switch. For details, see the
operation manual of the processor.
(1) When a scope switch is pressed, the function assigned to the switch is activated.
5.2.4 Bending Operation
WARNING
• Do not forcibly turn the angulation knob further after turning the knob until it stops. If the
angulation knob is forcibly turned, the angulation mechanism may malfunction and the
bending section may not return to its neutral position, making it difficult to withdraw the
endoscope.
• If the bending section does not return to its neutral position during a procedure, do not
withdraw the endoscope forcibly. Consult your local FUJIFILM dealer. If the endoscope
is withdrawn forcibly, injury to tissues in the body cavity, bleeding and/or perforation may
result.
• Be extremely careful when performing retroflexed observation in a narrow lumen. Do
not perform retroflexed observation forcibly. Otherwise, it may become impossible to
straighten the angle of the bending section and/or withdraw the endoscope from the
patient.
CAUTION
• Avoid forcible or excessive angulation as this may impose a severely heavy load on the
wire controlling the bending section. This may cause stretching or tearing of the wire.