AK 95 / AK 95 S Service Manual
HCEN 9134 Rev. 10.2005
Technical Description
2:3
Level Type of parameters Adjustable by...
4 Treatment parameters Operator, nurse
3 Alarm limits and Clinics technician
defaults
2 Clinic or customer Gambro technician or
adaptions Gambro subsidiary
1 Market adaptions Gambro Head office
0 Safety limits Fixed
System and Treatment
Parameters
The system and treatment parameters can be
set at five different levels (table 1).
Table 1. System and Treatment parameters
For information about the use and functions of
the parameters in level 2 and 3, see the chapter
Technicians Guide.
For information about the use and functions of
the parameters in level 4, see
AK 95 Operators Manual, HCEN 9542 or
AK 95 S Operators Manual, HCEN 9794.
Figure 1. AK 95 S.
A
B
Acetate
1002

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Gambro AK 95 S and is the answer not in the manual?

Gambro AK 95 S Specifications

General IconGeneral
TypeHemodialysis Machine
ManufacturerGambro
ModelAK 95 S
Dialysate Flow Rate300-800 ml/min
Power Supply100-240 V AC, 50/60 Hz
Arterial Pressure MonitoringYes
Venous Pressure MonitoringYes
Intended UseHemodialysis
ApplicationHemodialysis, Hemofiltration, Hemodiafiltration, Isolated Ultrafiltration
Blood Flow Rate50-500 mL/min

Related product manuals