HCEN12745 Revision 10.2016
Program version 2.xx
AK 98 Dialysis Machine - Attentions B:179
612 No patient ID is confirmed
612 No patient ID is confirmed
Confirm a patient ID. To continue without a
patient ID, press Confirm.
Appears:
When the machine is in pre-treatment mode or when
Connect patient is selected, and the patient“s ID is not
defined.
Machine actions:
None
Operator action:
1. Enter the patient“s id and press Confirm.
When it is not required to enter a patient ID:
1. Press the Confirm button.
Mute time:
Permanent