WRO 300 H
HCEN12481-02/2012
Contact information
(To be filled in by the hospital/clinic)
If you need assistance in setting up, using or maintaining the
WRO 300 H; or if you want to report unexpected operation or
events, please refer to:
Postal address:
Phone number and/or web contact address:

Other manuals for Gambro WRO 300 H

Questions and Answers

  • B
    Brandon VelezSep 23, 2025
    What to do if Gambro WRO 300 H shows Insert Wand?
    • S
      Sharon SandersSep 23, 2025
      If a Central Chemical disinfection has been initiated via the remote control and the Disinfection Wand is not inserted, first press the designated button to silence the buzzer. Insert the Disinfection Wand to continue the Central Chemical disinfection procedure.
  • P
    Patrick ColeAug 23, 2025
    What to do if Gambro WRO 300 H has a conductivity alarm?
    • D
      donald49Aug 23, 2025
      If the Gambro Water Filtration Systems triggers a conductivity alarm, first press the buzzer to silence it. Note down the conductivity value. Continue treatment after treatment. Inform responsible medical staff according to local procedures and notify technical personnel for further actions.
  • L
    Levi BaileyAug 25, 2025
    What to do if Gambro WRO 300 H Water Filtration Systems has a conductivity stop?
    • K
      Kelly TorresAug 25, 2025
      If the Gambro Water Filtration Systems initiates a conductivity stop, first press the buzzer to silence it. Inform responsible medical staff according to local procedures. The system can be restarted by pressing the designated button, but it will stop again if the problem persists. If the issue continues, discontinue dialysis and notify technical personnel.
  • G
    Gloria JordanAug 26, 2025
    What to do if there is a Conductivity Alarm in Gambro Water Filtration Systems?
    • C
      Cody GarciaAug 27, 2025
      If the conductivity value has exceeded the alarm limit for more than 30 consecutive seconds, take these steps: First, press the designated button to silence the buzzer. Then, note down the conductivity value. After treatment, inform the responsible medical staff according to local procedures and notify technical personnel for further actions.
  • N
    nunezcynthiaAug 27, 2025
    What to do if Gambro WRO 300 H Water Filtration Systems shows Conductivity Stop?
    • T
      tiffanyhorneAug 27, 2025
      If the conductivity value has exceeded the stop limit for more than 30 consecutive seconds, you should: Press the button to silence the buzzer. Inform responsible medical staff according to local procedures. The unit can be restarted by pressing the button, but it will stop again if the problem persists. If the problem continues, discontinue dialysis and notify technical personnel.
  • T
    Terrence GreerSep 17, 2025
    What to do if Gambro WRO 300 H Water Filtration Systems shows Insufficient Intake?
    • N
      Nicole AdamsSep 18, 2025
      If the product water conductivity is below the minimum intake conductivity limit, first press the designated button to silence the buzzer. If the Disinfection Wand is inserted and the button is pressed, a second intake phase is started. If the Disinfection Wand is not inserted and the button is pressed, the unit continues to the Dwell period.
  • C
    cameron68Sep 14, 2025
    What to do if Gambro Water Filtration Systems shows High Level Sensor Failure?
    • J
      Jason ThompsonSep 14, 2025
      If there is a failure of the high level sensor in the tank, first press the designated button to silence the buzzer. Then, notify technical personnel.
  • C
    Cynthia PattersonSep 15, 2025
    What to do if Gambro WRO 300 H Water Filtration Systems shows Remove Wand?
    • S
      Sandra KimSep 16, 2025
      If the wand connector is not removed after completion of Chemical intake, first press the designated button to silence the buzzer. Then, remove the wand from the disinfectant intake port.
  • E
    Erica DunlapSep 6, 2025
    What to do if Gambro Water Filtration Systems shows Incorrect Water Level?
    • I
      ibeasleySep 6, 2025
      If the water level in the tank is lower than expected, first press the designated button to silence the buzzer. Check that the dialysis machine is turned off. Press the button until the unit either goes to Standby or an Active Cooldown. If the problem still persists, notify technical personnel.
  • K
    Kevin TaylorSep 3, 2025
    What to do if Gambro Water Filtration Systems shows Low Rejection Rate?
    • E
      Evan Aguilar MDSep 3, 2025
      If the Gambro Water Filtration Systems indicates a low rejection rate due to decreased membrane performance, first press the designated button to silence the buzzer. Perform cleaning according to local procedures. If the problem persists, notify technical personnel.

Summary

Device Operation

Regular Maintenance Procedures

Heat and Low Flow Heat Disinfection

Guidelines and procedures for heat disinfection and low flow heat.

Chemical Disinfection and Related Procedures

Details disinfectants, procedures, residual testing, and Central Chem.

Cleaning Processes and Agents

Covers acid and alkaline cleaning, agents, and procedure steps.

Technical Data and Specifications

Troubleshooting and Error Handling

General and Operational Alarms

Details common alarms like sensor failures, internal errors, and conductivity issues.

Process-related Alarms

Covers alarms related to heat, water supply, chemical processes, and cleaning.

Specific Error Codes and Their Resolution

Addresses specific error codes for rinse, intake, temperature, and water level issues.

Gambro WRO 300 H Specifications

General IconGeneral
ApplicationHemodialysis water treatment
TypeReverse Osmosis System
ModelWRO 300 H
Power Supply230V, 50/60 Hz
Water QualityMeets AAMI Standards
Filtration CapacityUp to 300 liters/hour (at 15°C)
Temperature Range5 - 40°C