1 Technical data
1.1 JJ223BF-JJ1523BF (3 Digits)
Modell
JJ223BC
JJ323BC
JJ623BC
JJ1023BC
JJ1523BC
Weighing range (max)
220g
320g
620g
1020g
1520g
Readability ( =d)
1mg
1mg
1mg
1mg
1mg
Tare range
0-220g
0-320g
0-620g
0-1020g
0-1520g
Linearity
± 3mg
± 3mg
± 3mg
± 3mg
± 3mg
Reproducibility
± 2mg
± 2mg
± 2mg
± 2mg
± 2mg
Stabilization time
< 3 s
< 4 s
Operating
temperature
15-30 °C
18-23 °C
Humidity
(40-80)%RH
(50-70)%RH
Weighing plate
Ø 80 mm
Ø 120 mm
Calibration weight *
200g
200g
500g
500g
1000g
1000g
Minimum weight **
20 mg
Warming up time
1 h.
Minimum piece
weight ***
1mg
Reference piece
numbers for piece
count
1, 10, 20, 50, 100
Weighing units
g (gram), ct (carat), ozt (troy ounce), oz (ounce),
GN (gran), dwt (pennyweight), lb (pound)
Dimensions
(W x D x H)
220 (B) × 360 (T) × 345 (H) mm
Net weight
10 kg
Underfloor weighing
Yes
Interface
RS 232C
Power supply
AC 230V±10% 50Hz/60Hz
Auto power off
No
* Needed reference weight for a new calibration
**Minimum weight that must be placed on the scale. If you place 1mg on a scale with
20mg minimum weight, it will not be shown and stay on 0
*** Minimum weight of a single piece during counting