GANSHORN Medizin Electronic GmbH Service Manual Ref.No. 03 140 0 046
PowerCube-Ergo-CASE_ENG_rev.02, 23.11.2012 chapter 3, Disassembly, page 8of 10
_____________________________________________________________________
How to Remove / Replace the CO2-Analyzer
1) Remove the cover correctly from the base assembly
2) Remove the electronic unit correctly including removing the CO2-sensor cable
3) Flap out the compensation vessel to make the CO2-sensor accessible
4) Remove all tubes from the CO2-Analyzer
5) Unscrew the three attachment screws and withdraw the CO2-Analyser from the montage plate
To assemble the modules again perform the same steps backwards
Please take care to ensure the correct flow direction when connecting the tubing:
a. To the IN-socket of the CO2-Analyzer connect the thick tube coming from the M-gas-IN-
socket in the PowerCube’s rear panel
b. To the OUT-socket of the CO2-Analyzer connect the thick tube coming from
compensation vessel
After replacement of the CO2-Analyser please perform:
1) CO2-Offset adjustment (see chapter 6, Troubleshooting, checks and adjustments)
2) CO2-Gain adjustment
3) Gas calibration
Pict.: PowerCube-Ergo, inner lateral