Checklist for PowerCube Ergo /SSO Rev E
Revision: Rev01 30.11.2012
1. Examinee
This check is only to be carried out from authorized persons (MPBetreibV §11)!
All following parameters have to be checked under ambient conditions allowed for measurement use and with a
supply voltage of 230 V / 50Hz!
Respect the warm-up time of 30 minutes!
The pressure sensors must be checked with pos. und neg. pressure; the higher deviation applies!
2. Check ambient conditions
Temperature; tolerance from +15 °C to +35 °C
Rel. humidity ; tolerance from 30% to 80%
Ambient pressure; tolerance from 700 hPa to 1050 hPa
4. Test and calibration gas
Ergo; 15,9% O
2
4.5, 5% CO
2
4.5, Rest N
2
4.5
5. Deviations, Establishments and Evaluation