EasyManua.ls Logo

gc Labolight DUO - User Manual

gc Labolight DUO
176 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
Instructions for Use
Gebrauchsanweisung
Mode d'emploi
Istruzioni per l'uso
Instrucciones de uso
Instrukcja obsługi
Labolight DUO
EN
LED Dual Mode Light Curing Unit for Indirect Composite Resin
DE
Lichthärtungs-Gerät mit dualer LED-Lichtquelle für indirekte Komposite
FR
Unité lumineuse LED à mode dual pour la photopolymérisation des résines composites indirectes
IT
Lampada LED a foto-polimerizzazione duale per resine composite da restauro indiretto
ES
Unidad de Fotopolimerización LED dual para resinas de composite indirecta
PL
Dwumodułowa lampa LED do polimeryzacji świetlnej żywic kompozytowych w technice pośredniej

Table of Contents

Questions and Answers

  • B
    Benjamin SalazarAug 12, 2025
    How to troubleshoot a gc Lighting Equipment that won't power on?
    • J
      Jackson MccarthyAug 13, 2025
      If the main switch is on but the display remains blank, ensure that all cords (main body, AC adapter, power cord, and outlet) are securely connected. Verify that you are using the dedicated power cord and AC adapter that came with the gc Lighting Equipment. A defective AC adapter or power cord may also be the reason, or the fuse might have blown. In these cases, replacements are needed.
  • R
    Robyn WhitakerAug 16, 2025
    What to do if the overheat indicator is lit and the remaining time blinks on my gc Labolight DUO?
    • J
      Juan BrownAug 16, 2025
      If the overheat indicator is lit on your gc Lighting Equipment and the remaining time is blinking, it means the LEDs have temporarily overheated and irradiation is temporarily halted. Allow the LEDs to cool down before restarting. If this happens frequently, further assistance may be required.
  • N
    nicholasthomasAug 20, 2025
    What does it mean when the overheat indicator is lit and 'Fin' is displayed on my gc Labolight DUO?
    • B
      Bethany RayAug 20, 2025
      If the overheat indicator is lit and 'Fin' is displayed on your gc Lighting Equipment, it indicates a temporary halt due to the LEDs overheating after irradiation is completed. Allow the LEDs to cool down before restarting. If this issue persists, further assistance may be required.
  • S
    scottwoodAug 23, 2025
    Why is the overheat indicator lit on my gc Labolight DUO?
    • S
      Scott JenningsAug 23, 2025
      If the overheat indicator is lit on your gc Lighting Equipment, it signifies a temporary halt because the LEDs are too hot. Wait for them to cool down before attempting to restart the device. If this issue occurs frequently, further assistance may be required.
  • S
    steve24Aug 27, 2025
    What to do if the door indicator is lit and the set time doesn't count down on my gc Labolight DUO?
    • K
      kallenAug 27, 2025
      If the door indicator is lit when you start irradiation in full mode on your gc Lighting Equipment and the set time isn't counting down, it's likely that the door isn't fully closed. Check for obstructions in the door's path and ensure it is completely closed before using the device.
  • I
    Isabel DavisAug 31, 2025
    Why are the upper row LEDs unlit on my gc Lighting Equipment?
    • A
      Amber IrwinAug 31, 2025
      If the display on your gc Lighting Equipment is normal but the upper row LEDs are unlit, it may indicate a failure or disconnection of the LEDs. Further assistance may be required.
  • A
    Anthony HarrisonSep 4, 2025
    Why are all the LEDs lit on my gc Lighting Equipment?
    • D
      David PatrickSep 4, 2025
      If the display on your gc Lighting Equipment is normal and all LEDs are lit, it could be due to contamination or scratches on the optical components. Try cleaning inside the curing chamber and lens cover. If necessary, replace the lens cover.
  • A
    anna15Sep 8, 2025
    Why are no LEDs lit on my gc Labolight DUO Lighting Equipment, even though the display is normal?
    • D
      Deborah TaylorSep 8, 2025
      If the display on your gc Lighting Equipment is normal, but no LEDs are lit, it suggests a failure or disconnection of the LEDs. Further assistance may be required.

gc Labolight DUO Specifications

General IconGeneral
Power SupplyAC 100-240V, 50/60Hz
Light SourceLED
Dimensions100 mm
Weight1.5 kg
Timer Settings5 seconds
Irradiance2000 mW/cm²
TimerYes