Do you have a question about the GE Druck DPI 150 and is the answer not in the manual?
Manufacturer designed equipment for safe operation as detailed in this manual. User must not use for other purposes.
Do not apply pressure greater than the maximum working pressure stated in the specification.
Protect set-up menus; unauthorized access degrades performance and settings. Factory PIN is 0268.
Lists safety and EMC standards (EN61010, EN61326) the product complies with.
Lists available gauge pressure ranges in mbar and bar.
Specifies the maximum working pressure for different ranges.
Details the precision, including non-linearity, hysteresis, repeatability, and temperature effects.
Specifies the accuracy of the instrument, including pressure measurement uncertainty.
Explains the function and comments for each front panel key.
Setting the upper limit for calibration or span adjustment.
Setting the lower limit for calibration or span adjustment.
Sets the duration for the airspeed leak test.
Specifies the airspeed range for the leak test.
Allows selection of units for airspeed measurement.
Explains the tare function for airspeed measurements.
The Druck DPI 150 is a high-accuracy, single-range pressure indicator designed to provide precise pressure readings in both pressure measurement and aeronautical units. It utilizes a Druck IDOS sensor for its core functionality. The instrument is housed in a molded plastic case with integrated rubber feet, making it suitable for workbench use.
The primary function of the DPI 150 is to display pressure readings. It can operate in various modes, including a standard indicator mode, pressure leak test, airspeed leak test, airspeed measurement, and Mach number calculation.
In Indicator mode, the device displays the current pressure reading in selected units. Users can configure the display to show full-scale values, apply filters, and set tare values. The tare function allows the user to zero the displayed value and store the difference (offset) in the pressure sensor's non-volatile memory, which is useful for relative pressure measurements. The device supports a wide range of pressure units, including mbar, bar, Pa, hPa, kPa, MPa, atm, psi, mmHg, mHg, mH₂O, inHg, inH₂O, kgcm², kgm², inH₂O4, ftH₂O, ftH₂O4, and inH₂O60. It also offers options for setting a time constant and band for filtering the displayed readings, which can be useful for stabilizing fluctuating pressure signals. The indicator mode also includes features for setting high and low span values, which are likely used for scaling or alarm functions.
The Pressure Leak Test mode is designed to detect leaks in a pressure system. Users can set a test time, and the device will monitor the pressure change over that period, displaying the start pressure, end pressure, pressure change, and leak rate. This is crucial for ensuring the integrity of pneumatic systems.
Similarly, the Airspeed Leak Test mode, when correctly enabled, performs a leak test specifically for airspeed systems. It functions much like the pressure leak test, allowing users to set a test time and displaying relevant parameters such as start pressure, end pressure, pressure change, and leak rate, but in airspeed units like kts, mph, or km/h.
The Airspeed mode, also requiring correct enablement, directly measures and displays airspeed in kts, mph, or km/h. This mode also includes options for filtering, taring, and setting time constants and bands for signal conditioning.
The Mach mode, when enabled, calculates and displays the Mach number based on the measured pressure and a set reference pressure. Users can set the reference pressure, which is essential for accurate Mach number calculations.
The DPI 150 features a user-friendly front panel with function keys for easy navigation and operation.
esc key: Steps back one selection without changing the current setting.menu OK key: Steps back one selection, sets and saves the value, and selects and enters menus.hold key: Freezes the displayed value while the instrument continues measuring. Pressing the key again resumes live display.zero key: Changes the displayed value to 0.000 and stores the offset in the pressure sensor's non-volatile memory.The instrument offers several set-up menus for customization:
The DPI 150 supports various connections on its rear panel:
The device can be used as a bench stand by lifting its front to provide a better display angle and easier key-pad access. It can also be panel mounted using an optional kit, which requires removing the rubber feet and securing side and front plates into a panel cut-out. Adequate air circulation is important for panel-mounted installations to cool the instrument.
The manual emphasizes the importance of safety during operation and maintenance. Users are advised to switch off the power supply and depressurize pipes before connecting or disconnecting the instrument. Only suitably qualified personnel should perform procedures outlined in the manual.
The Status menu provides important information about the device:
The supervisor and calibration menus are PIN-protected to prevent unauthorized changes that could degrade performance or accuracy. If using a power adaptor not supplied with the instrument, it must comply with SELV safety requirements of EN61010. For technical advice or service, users are directed to contact the manufacturer or subsidiary.
| Protection Rating | IP54 |
|---|---|
| Input Method | Touchscreen |
| Operating Temperature | -10°C to +50°C |
| Pressure Media | Non-corrosive gases and liquids |
| Power | Rechargeable battery or external power supply |
| Communication | USB |
| Pressure Units | bar, psi, kPa |
| Storage Temperature | -20°C to 70°C |