Page #64 background image
Site Preparations
3-12 Ultrasound SystemCommon Service Information
Direction 5444964-100 English
Rev. 5
Electrical requirements for Laptop Ultrasound Systems (continued)
Table 3-8: Electrical Requirements - Vivid Laptops
Product Voltage Tolerance
Current or
Power
Consumption
Frequency
AC/DC Converter for
Vivid i/Vivid iN and
Vivid q/ Vivid qN
100 - 240 VAC ±10% 0.5 to 1.0 A 50-60 Hz
Adapter for Vivid e 100 - 240 VAC ±10% 130 VA max. 1.3 A
(max.)
50/60 Hz
Table 3-9: Electrical Requirements - Voluson Laptops
Product Voltage Tolerance
Current or
Power
Consumption Frequency
Voluson i/e 100 - 240 VAC
±10%
120 VA
50, 60 Hz
(±2%)
Voluson Station /
Voluson Dock Cart
100-130 VAC
220-240 VAC 320 VA

Table of Contents