EasyManuals Logo

GE LOGIC C2 Series User Manual

GE LOGIC C2 Series
127 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #66 background imageLoading...
Page #66 background image
Site Preparations
3-14 Ultrasound SystemCommon Service Information
Direction 5444964-100 English
Rev. 5
Electrical Requirements for Isolation Cart
Table 3-11: Isolation Cart for LOGIQBook XP Series/LOGIQ e/LOGIQ e Vet/LOGIQ i/
Vivid e
Input Output
100-120V~, 50/60Hz, 500VA 100-120V~, 50/60Hz, 250VA
220-240V~, 50/60Hz, 500VA 220--240V~, 50/60Hz, 250VA
Table 3-12: E-Isolation Cart for LOGIQ e
Input Output
100-120V~, 60Hz, 400VA 100-120V~, 60Hz, 350VA
220-240V~, 50Hz, 400VA 220--240V~, 50Hz, 350VA
Table 3-13: Advanced Isolation Cart
Input Output
100-120V~, 60Hz, 400VA 100-120V~, 60Hz, 350VA
220-240V~, 50Hz, 400VA 220--240V~, 50Hz, 350VA