EasyManua.ls Logo

GE LOGIQ A5 User Manual

GE LOGIQ A5
127 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Page #1 background imageLoading...
Page #1 background image
Technical Publication
Ultrasound System
Common Service Information
Direction Number: Direction 5444964-100 English
Rev. 5
BASIC SERVICE DOCUMENTATION.
COPYRIGHT GENERAL ELECTRIC COMPANY.
© 2013-2018 General Electric Company. All Rights Reserved.

Table of Contents

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ A5 and is the answer not in the manual?

GE LOGIQ A5 Specifications

General IconGeneral
Power Supply100-240 V, 50/60 Hz
Imaging ModesB-mode, M-mode, Color Doppler, Power Doppler, Tissue Harmonic Imaging
Transducer OptionsConvex, Linear, Phased Array, Endocavitary
ApplicationsAbdominal, Vascular, Musculoskeletal, Urology
DICOM CompatibilityYes
Display15-inch LED-backlit color LCD

Summary

Chapter 1 — Overview

Product description

Describes the ultrasound system's digital architecture, scanning modes, and signal flow.

Products Covered in this Manual

Lists all specific GE ultrasound models and configurations included in this CSI manual.

Chapter 2 — Safety Information

Important conventions

Explains document conventions, safety precaution messages (DANGER, WARNING, CAUTION), and hazard icons.

Safety considerations

Details critical safety precautions for human, mechanical, and electrical aspects during operation and servicing.

Dangerous procedure warnings

Highlights warnings for procedures carrying significant risks during system operation, emphasizing adherence to instructions.

Lockout/Tagout (LOTO)

Outlines essential requirements and procedures for safe energy control and lockout during servicing to prevent accidental energization.

Electromagnetic compatibility (EMC)

Discusses EMC principles, compliance standards, and electrostatic discharge (ESD) prevention measures for system integrity.

Chapter 3 — Site Preparations

Environmental specifications for Ultrasound System system

Details cooling, lighting, and environmental conditions required for optimal system setup and performance.

Electrical requirements

Specifies dedicated power, grounding, voltage, frequency, and outlet needs for safe and proper system installation.

EMI limitations

Explains susceptibility to EMI, potential sources, and prevention/abatement strategies to ensure signal integrity and performance.

Facility needs

Outlines purchaser responsibilities, required facility preparations, and desirable features for successful system installation.

Chapter 4 — Care and Maintenance

System maintenance

Covers preliminary checks, functional checks, peripheral/option checks, and physical inspection procedures for system upkeep.

Electrical safety tests

Details critical electrical safety tests including UPS, leakage currents, grounding continuity, and outlet wiring verification.

Leakage current limits

Provides specific acceptance limits for Earth, Chassis, and Applied Part leakage currents based on relevant IEC safety standards.

When there's too much leakage current …

Addresses troubleshooting steps for common leakage issues like AC/DC fails, chassis fails, probe fails, and others.

Ultrasound Equipment Quality Check (EQC and IQC)

Guides on performing Equipment and Image Quality Checks, typically coordinated with a GE Service Representative.