EasyManua.ls Logo

GE Vivid E9 - Page 62

GE Vivid E9
80 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
(RV Volume Acquisition – cont.)
7. Press Multi beat*
Tips: Adjust frame rate knob for optimal balance between
endocardial border visualization (spatial resolution) and
endocardial motion (temporal resolution)
Use Angle button on control panel – SAX volume view –
allows “centering” of the RV in the capturing volume
See 4D Acquisition quickcard for more Multi beat 4D
acquisition steps & tips
* Most RV Volumes require Multi beat acquisition for higher frame rates in
order to use the RV Vol tool accurately. If using single beat skip steps 6&7
6
7
RV Volume Acquisition
(RV Volume Acquisition – cont.)
8. Press Image Store or Freeze
Tip: If ‘‘preview on’’ – Use rotary knob for Cycle
Select to pick desired loop. If Freeze is pressed
then press Cine to review loop options. When
desired cycle loop is playing – Press Image Store
again to store
9. Recall image, and adjust UD Clarity
for smoothness
10. Use 9 Slice to check for volume completeness, image quality, and artifacts
Tip: 9 Slice should show continuous edges for each crescent-shaped short
axis (sagittal) view, from apex to TV annulus
10
9

Other manuals for GE Vivid E9