Site Preparations
3-16 Ultrasound SystemCommon Service Information
Direction 5444964-100 English
Rev. 5
Electrical requirements for the Ultrasound workstations
Contents:
• ‘Electrical Requirements - EchoPAC PC Turnkey
Workstations’ on page 3-16
• ‘Electrical Requirements - LOGIQworks Workstations’ on
page 3-16
Table 3-15: Electrical Requirements - EchoPAC PC Turnkey Workstations
Product Voltage Tolerance
Current or
Power
Consumption Frequency
All models
100 VAC
±10%
8 A
50-60 Hz
110 VAC 8 A
230 VAC 4 A
Table 3-16: Electrical Requirements - LOGIQworks Workstations
Product Voltage Tolerance
Current or
Power
Consumption Frequency
All models 110 - 230 VAC ±10% 310 to 543 W
(depending on
workstation
model)
50-60 Hz

Table of Contents

Other manuals for GE Vivid i

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Vivid i and is the answer not in the manual?

GE Vivid i Specifications

General IconGeneral
TypePortable Ultrasound System
Transducer Ports2 active ports
Doppler ModesColor Doppler, Power Doppler, Pulsed Wave Doppler, Continuous Wave Doppler
Imaging Modes2D, M-Mode
ConnectivityDICOM, Ethernet, USB
ApplicationsVascular, Abdominal
WeightApproximately 5.5 kg (12.1 lbs)
Battery LifeUp to 1 hour (depending on usage)
ProbesPhased array, Linear array, Convex array

Related product manuals