EasyManua.ls Logo

General RHMG650 - Checking Calibration

General RHMG650
16 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
If the primary readout shows
0.0%
with the meter in your hand (and not touching anything), the
meter is ready to make measurements. To measure the moisture level of a material, press the
pinless sensor on the back of the meter (Fig. 1, Callout L) against it. The moisture level, as
percentage, will appear on the primary readout.
If the primary readout shows a value other than 0.0%, you should reset the baseline of
measurements in
MOIST INT
mode. To do so, continue to hold the meter in your hand (touching
nothing) and press the
ZERO
button. Immediately after the term
ZERO
appears briefly below the
bar graph at the left of the display, the primary readout will show
0.0%
and the measurements
will again be accurate in
MOIST INT
mode.
To obtain accurate readings in
MOIST INT
mode, you should reset
the readout to 0.0% each time you re-enter that mode from
another mode.
CHECKING CALIBRATION
Although you cannot calibrate the RHMG650 yourself, you can
check
the meter’s calibration
periodically to assure yourself of the accuracy of humidity-related readings (RH, absolute
humidity, dew point and mixing ratio). Factory calibration should minimize the humidity sensor’s
drift for several years—well beyond the warranty period.
At a minimum, a calibration check should entail immersing the slotted structure protecting the
humidity sensor in two bottles containing saturated salts. One reference salt should produce a
reading of a 33%; the other should produce a reading of 75%.
Two salts of this type are available from General: the HR33 (33%) MgCl Calibration Salt and the
HR75 (75%) NaCl Calibration Salt.
Depending on the size and shape of the calibration salt bottle that you use, you may have to
remove the hanger tab (Fig. 1, Callout B) from the meter to provide sufficient clearance for the
bottle to fit snugly over the slotted structure protecting the humidity/temperature sensor (after
removing the dust cap). The hanger tab is removable for that reason.