1002 - 1013 Wasserbäder
Water Baths
Please make sure that no fluids come into
contact with cable connections or the
electrical part of the inside unit!
Always pull the plug before opening or
cleaning the unit due to the danger of
electrical shocks.
For cleaning purposes, the perforated tray can be taken out of the bath.
The inside bath can be descaled with suitable descaling agents (eg "tin-be" by Messrs. Dr.
Otto Hartmann, 71665 Vaihingen/Enz, Germany).
Do not use any hydrochlorid acid product!
These products damage the unit's interior and screw connections!
The polished exterior of the unit can easily be kept and restored with usual stainless steel
polishing solutions (eg "Helios" by Messrs. Henkel Hygiene GmbH, 40589 Düsseldorf,
Germany). The off-white powder coated parts can be cleaned with mild detergents if
necessary.
It is advisable to regularly exchange the water within the bath. The water outlet tap is situated
at the back of the unit.
The GFL Water Bath was set and calibrated at a temperature of 50°C. For re-calibration
during servicing, a calibration manual can be obtained from GFL on request. Please enquire
by fax and mention type and serial number of the unit in question.
Servicings, repairs or modifications must be carried out according to the commonly
recognised Technical Rules and Regulations by competent electricians only.
Only original spare parts must be used. Always demand a detailed confirmation of the carried
out tasks by the person in charge (company, date, signature).
Seite 13 Rev.03