1083 / 1086 Schüttelwasserbad
Shaking Water Bath
12
The shaking device is operated by the switch in the starter panel. The shaking frequency
(reciprocating) can be increased by turning the relevant knob clockwise. It can be set in a
range of 10 to 250 min
-1
.
Shaking Water Bath 1086 disposes of an LED display for the actual shaking frequency.
11. Servicing and Maintenance
GFL Shaking Water Baths are produced with first class materials and are made to withstand
even rough service conditions. Nevertheless, the units should only be subjected to rough
conditions within sensible limits.
Please make sure that no fluids come into
contact with cable connections or the
electrical part of the inside unit! Always
withdraw plug before opening or cleaning
the unit due to the danger of electrical
shocks.
For cleaning purposes, the shaking device can be taken out of the bath. For this purpose, the
device has to be pushed to the max right and unscrewed via the knurled screw (connection
to motor, Pos. 4 in the exploded view). The device can now be pushed to the far left and
carefully lifted out of the bath.
The inside bath can be descaled with suitable descaling agents (eg "tin-be" by Messrs. Dr.
Otto Hartmann, 71665 Vaihingen/Enz, Germany).
Do not use any hydrochlorid acid product!
These products damage the unit's interior and screwings!
The polished exterior of the unit can easily be kept and restored with usual stainless steel
polishing solutions (eg "Helios" by Messrs. Henkel Hygiene GmbH, 40589 Düsseldorf,
Germany). The off-white powder coated parts can be cleaned with mild detergents if
necessary.
It is advisable to regularly exchange the water within the bath. The water outlet tap is situated
at the back of the unit.
The Shaking Water Bath was set and calibrated at a temperature of 50°C. For re-calibration
during servicing, a calibration manual can be obtained from GFL on request. Please enquire
by fax and mention type and serial number of the unit in question.
Servicings, repairs or modifications must be carried out according to the commonly
recognised Technical Rules and Regulations by competent electricians only.
Only original spare parts must be used. Always demand a detailed confirmation of the carried
out tasks by the person in charge (company, date, signature).