5. Operation
A mechanical clockwork serves to run the Shaker in continuous or timed operation.
All models are equipped with ac motors, protected against overload. The electronic frequency
control ensures a gentle start-up, irrespective of load.
A stable and durable mechanical construction serves to impart the torque of the motor with the
specific shaking motion to the shaking platform.
6. Servicing, maintenance and clearing possible defects
Please make sure that no fluids come into contact with cable
connections or the electrical parts of the inside unit!
Before cleaning and/or opening the unit (only by an electrician!)
always pull the plug from the mains socket.
Danger of electrical shock !
Exchanging the fuses
The two fuses of the shaker are situated in a drawer below the unit plug. In order to check or to
exchange these fuses, press the fixing clips on both sides of the drawer and pull.
The fuses may only be exchanged against fuses of the same type. Information on the fuse
type can be found on the nameplate next to the unit plug and in the spare parts list of
these operating instructions.
Breakdown in case of overload and mains failure
Overheating of the shaker due to overload will cause the shaker to be switched off.
Caution, the unit will restart automatically after cooling down or after mains failures.
If the unit is switched off due to any kind of breakdown, it must by all means be switched off,
before the shaker is touched.
GFL Shakers are produced with first class materials and are made to withstand even rough
service conditions. Nevertheless, the units should only be subjected to rough conditions within
sensible limits.
The off-white pulver-coated surfaces of the housing and the shaking platform may be cleaned
with mild detergents, if necessary.
Please do not hesitate to contact your local dealer or us for technical support.
In case of complaints please contact your local dealer or us:
GFL Gesellschaft für Labortechnik mbH
Schulze-Delitzsch-Strasse 4 D-30938 Burgwedel Germany
Servicings, repairs or modifications must be carried out according to the
commonly recognised Technical Rules and Regulations by competent electricians
only.
Only original spare parts must be used. Always demand a detailed confirmation of the carried out
tasks by the person in charge (company, date, signature).