Questions and Answers

  • B
    Bryce JohnsonSep 23, 2025
    Why is the temperature less than desired on my GGM VH-2600-115 Humidifier?
    • J
      Jenna BoltonSep 23, 2025
      If the temperature is less than desired on your GGM Humidifier and there is no alarm, the humidifier may be in the process of warming up. Wait for a few minutes to see if the temperature reaches the desired level.
  • L
    Lisa WolfAug 25, 2025
    Why is there excessive water in the inspiratory circuit of my GGM VH-2600-115 Humidifier?
    • T
      Travis SawyerAug 25, 2025
      Excessive water in the inspiratory circuit of the GGM Humidifier, also known as rainout, can occur for several reasons. If in non-heated wire mode, reposition the water trap to the lowest point of the circuit. If in heated wire mode, increase the distance from the heater wire to the temperature probe so that the distance is between 25 mm and 100mm, ideally 30mm. If the inspiratory circuit is being cooled by an air conditioner, shield the circuit or use a heater wire circuit. If the chamber control is set too high, reduce the setting. If the maximum permissible peak flow rate of the humidification chamber has been exceeded, use the correct chamber referring to the operating instructions for the chamber type. Ensure the chamber is level. Finally, check the temperature probe for any faults.
  • B
    Brandon RiveraAug 25, 2025
    How to fix airway temperature low alarm on GGM Humidifier?
    • K
      Kristen LucasAug 26, 2025
      If the airway temperature low alarm is activated on your GGM Humidifier, first check the position of the temperature probe in the circuit. If the gas flow has stopped or been interrupted, turn the humidifier off until gas flow resumes. Ensure the temperature probe is correctly fitted in its housing. Also, ensure the gas flow rate is more than 5 liters per minute; increase it if it's too low.
  • V
    Victoria Taylor MDAug 27, 2025
    What to do if the chamber temperature high alarm is activated on GGM VH-2600-115?
    • N
      Nicholas McfarlandAug 27, 2025
      If the chamber temperature high alarm is activated on your GGM Humidifier, first refill the humidification chamber if there is no water in it. Also, check the chamber sensor on the temperature probe for faults.
  • A
    Andrew HinesAug 27, 2025
    How to resolve airway temperature high alarm on GGM Humidifier?
    • S
      Seth CrossAug 27, 2025
      If the airway temperature high alarm is activated on your GGM Humidifier, wait for the temperature to fall below the threshold if there was a sudden spurious increase in gas flow. If the heater wire is too close to the temperature probe, adjust the wire or hose so that the distance is between 25 mm and 100 mm, ideally 30 mm. Check the airway sensor on the temperature probe for faults.
  • R
    Robert Romero Jr.Sep 15, 2025
    What does ‘ER.03’ mean on my GGM VH-2600-115 display?
    • R
      Ryan KlineSep 15, 2025
      If you see ‘ER.03’ on your GGM Humidifier display, it indicates the heater plate is broken or the heater plate overheated protector has been activated. You can try to reset the overheated protector. If that doesn't work, the unit may need repair.
  • R
    reillyjonathanSep 12, 2025
    What to do if the expiratory heater wire alarm goes off on my GGM VH-2600-115?
    • P
      Patricia LynchSep 13, 2025
      If the expiratory heater wire alarm activates on your GGM Humidifier while the temperature is still displayed, the expiratory heater wire may be broken. Replace the heater wire, or push the HEATER WIRE button twice.
  • K
    Kelsey DyerSep 14, 2025
    How to troubleshoot a breathing circuit leak in GGM VH-2600-115 Humidifier?
    • E
      Ellen WilliamsSep 14, 2025
      If the GGM Humidifier has a breathing circuit leak, check all circuit connections. Ensure the water bag is connected. Also, check that the temperature probes are correctly fitted and connected.
  • W
    watsonamandaSep 5, 2025
    What does temperature probe alarm ‘C.SHT’ mean on the GGM VH-2600-115?
    • C
      Chelsea BellSep 5, 2025
      If your GGM Humidifier displays a temperature probe alarm ‘C.SHT’, it means the chamber temperature probe is short. Replace the probe.
  • A
    Alyssa StoneSep 3, 2025
    What does temperature probe alarm ‘A.SHT’ mean on my GGM VH-2600-115 Humidifier?
    • C
      Christopher CannonSep 3, 2025
      If your GGM Humidifier displays a temperature probe alarm ‘A.SHT’, it means the airway temperature probe is short. Replace the probe.

Summary

Product Specifications

2.3.2. Alarm Parameters

Outlines the conditions and triggers for various audible and visual alarms.

2.4. Performance

Covers recommended ambient temperature, flow range, and humidity output.

Explanation of Controls and Indicators

3.3. Front Panel

Details buttons, indicators, and display on the front panel.

3.4. Display

Explains symbols and codes displayed on the unit.

Maintenance Schedule

4.1. Maintenance Schedule

Outlines the necessity and frequency of regular maintenance.

4.1.5. Safety

Explains earthing resistance and protection against electric shock hazards.

Alarm and Troubleshooting

5.1. Alarm Indicator

Lists the types of alarms and their corresponding indicator lights.

5.2. Troubleshooting

Provides guidance on diagnosing and resolving common issues.

5.2.1. Low Temperature

Details causes and actions for low temperature alarms.

5.2.2. High Temperature

Details causes and actions for high temperature alarms.

5.2.3. Temperature Fluctuating

Explains causes for fluctuating temperatures and corrective actions.

5.2.6. Alarms

Details specific error codes and their explanations for alarms.

Servicing Procedures

6.1. General Precautions

Important warnings and procedures before performing maintenance.

6.2.2. Replacing Fuses

Instructions for identifying and replacing the unit's fuses.

6.2.5. Replacement of Printed Circuit Boards (PCBs)

Guide for replacing internal circuit boards.

Performance Testing

Recommended Maintenance Checklist

GGM VH-2600-115 Specifications