4
CUSTOMER SERVICE | FC 203B FRACTION COLLECTOR
CustOmer serviCe
Collection Windows
Collection windows allow for collecting only the effluent
desired and discarding the rest. Optionally, build up to ten
collection windows.
Collection windows are useful for:
• excluding void volumes
• excluding equilibration or washing volumes
• aiding in selective peak collection
Remember four key points about collection windows:
• It is always possible to collect from start to end,
designating the entire collection period as one
collection window.
• Whenever the fraction collector is outside a collection
window, it automatically diverts the effluent to drain.
• At the start of each collection window, the dispense
head automatically advances to the next tube.
• The collection windows are not mode specific.
Windows created in one mode are active in all other
modes.