5
ENGLISH
Lamp Replacement Procedure
Before removing the lamp, make sure that the device has been turned off for some minutes other wise you
run the risk of being burnt. Lamps may be hot. So it’s necessary to let it cool.
Remove Dermatoscope head from the power source as advised in
operating instructions (See g.1).
Pull out lamp with nail le or similar tool (if necessary) by getting it under
the base of lamp (See g.3).
Use recommended lamp for replacement by carefully aligning pin on lamp
shaft with slot in Dermatoscope head. Push the lamp until seated well.
Clean the contact plate with cleaning solution prior to
autoclaving.
Battery Replacement Procedure
In case of necessity replacement of new batteries then carefully handle
the batteries as the liquid in a leaking battery can irritate skin and eyes.
While replacing batteries it is necessary to use exclusive
alkaline type batteries.
a) Open the battery compartment, to do this, turn the end cap counter
clock wise.
b) Once opened, insert the batteries paying attention to the direction of
poles (See g 1).
c) To close the battery compartment, screw down the cap clock wise and
check that the batteries are in contact with poles.
Storage
Since the product is made of corrosion-proof materials suitable for environmental conditions foreseen for its normal
use, it does not require special care. However it is necessary to store it in a closed place making sure that is protected
from dust and dirt to assure it’s hygienic properties. Store in clean environment and preserve in normal temperature.
Precautions
Dermatoscope is equipped with halogen lamp and shall be used exclusively with alkaline batteries type “C”
correctly installed.
This shall be used by qualied medical personnel only.
Don’t use the equipment in case it is damaged. Apply to your retailer.
If the instrument has rusted then don’t use it.
Avoid precarious repairs.
Repairs shall be carried out with original spare parts only, which shall be installed according to the intended use.
Protect lamp surface against abrasions and scratches.
Disposal: The product must not be disposed of along with other domestic waste. The users must dispose of
this equipment by bringing it to a specic recycling point for electric and electronic equipment.
GIMA WARRANTY TERMS