34
9. The monitor can maintain the safety and performance characteristics for a
minimum of 10,000 measurements or three years, and the cuff can maintain
the performance characteristics for a minimum of 1000 measurements.
10. It is recommended the cuff should be disinfected 2 times every week if
needed (For example, in hospital or in clinique). Wipe the inner side (the
side contacts skin) of the cuff by a soft cloth squeezed after moistened with
Ethyl alcohol (75-90%), then dry the cuff by airing.
11. The monitor requires 6 hours to warm from the minimum storage tempera-
ture between uses until the monitor is ready for its INTENDED USE when
the ambient temperature is 20°C.
12. The monitor requires 6 hours to cool from the maximum storage tempera-
ture between uses until the monitor is ready for its INTENDED USE when
the ambient temperature is 20°C.
13. No servicing/maintenance while the monitor is in use.
ENGLISH