8
8. Specifications
Notice: The sensitivity should be dependent on the difference between the thermal
conductivities of gas and air (24.1(mW / m
㨯M
at 0 °C, 1atm); a large difference
results in a high sensitivity. Also, the kinds of gas and humidity should influence
the sensitivity.
Thermal conductivity of some common gases [(mW / m
k) at 0°C, 1atm]
The lithium-ion battery inside the LD239 does not apply battery regulations in Russia or
Australia. Do not bring the LD239 these countries.
Leak Detector LD239 Instruction Manual (Cat.No.2702-19336)
Rev.1-1
Printed in Japan Copyright © 2012 GL Sciences Inc.
Detector:
Charging micro USB cable(1m), Instruction Manual,
LD239 Battery(built-in), Sample Filter(built-in), Reference Filter(built-in)
Power supply:
Display:
Thermal conductivity of gases:
Sensitivity:
Overall size:
Accessories:
Operating temperature range:
Net weight:
Parameter setting:
Thermistor-actuated thermal conductivity cell
He, CO
2
, Ar, Ne, etc. (not available under flammable or corrosive atmosphere)
Selectable from 2 ranges: Standard(Std.) Range and High(Hi.) Range
[Std Range] Helium 0.005mL/min in atmosphere
[Hi Range] Helium 0.0005mL/min in atmosphere
[Leakage display] LCD display(Max. 8 bar graphs), LED(Max. 8 lamps)
[Parameter setting display] LCD display
Range(Std./Hi.), Buzzer(ON/OFF), LED(ON/OFF), LCD backlight(ON/OFF),
Auto Power Off(ON/OFF)
Rechargeable battery-powered equipment(built-in dedicated lithium-ion battery)
[Charging time] Max. 2.5 hours
[Continuous operation] Approximately 5 hours (with new, fully charged batteries)
(Std. Range, Buzzer OFF, LED OFF, LCD Backlight OFF)
10 - 40 °C(non condensing)
50(W) × 19.5(D) × 111(H) mm (excluding projection)
Approximately 95g
Hydrogen *
Helium *
Neon *
Methane *
168.2
142.2
46.5
30.2
Air
Nitrogen
Ethane
Ethylene
24.1
24.0
18.0
16.4
Argon
Carbon dioxide
Krypton
Xenon
16.3
14.5
8.7
5.2
Gas
thermal
conductivity
Gas Gas
thermal
conductivity
thermal
conductivity
Action for Environment (WEEE)
To users of GL Sciences equipment in the European Union:
The WEEE (Waste Electrical and Electronic Equipment) symbol on the product
indicates that it should be disposed of as agreed between the end user and the
distributor in a manner consistent with required WEEE Directives. It should not
be disposed of with general household waste.
: Gases which have higher thermal conductivity than air.
*
: Undetectable gases