6
Low battery. Change the batteries according to
the instructions (§7.3).
The temperature was not within the recom-
mended range. Repeat the measurement after at
least 30 minutes, when the operating tempera-
ture has been restored.
8.3. Unexpected test results
If the test result is unusual compared to your previous test results:
1. Repeat the measurement using a new test strip (§4.2).
2. Perform a control measurement using the GlucoMen® areo Control
solution (§4.3).
3. Change the test strip vial and repeat the measurement using a new
test strip (§4.2).
4. If you are still uncertain about the test results, contact your doctor or
healthcare professional.
WARNING
In case of an unexpected reading, consult your doctor or healthcare
professional before altering your treatment.
If the test results continue to be unusual, please, contact the
A. Menarini Diagnostics customer service.
9. Technical specifications
Product: GlucoMen® areo
Manufacturer: A. Menarini Diagnostics srl
Tested item: Blood Glucose level
Unit of
measurement:
mmol/L
Test Range: 1.1 – 33.3 mmol/L
Haematocrit range:
10 – 70 % (haematocrit compensated)
Test strip: GlucoMen® areo Sensor
Test strip coding: No coding required
Sample: Fresh capillary whole blood
Sample size: Minimum 0.5 µL
Test time: 5 seconds
Assay Method: Electrochemical, Glucose Oxidase (GOD,
Aspergillus Niger sourced) based method.
Mediator: Hexacyanoferrate (III) ion.
Calibration and
traceability:
The results are equivalent to the plasma glucose
concentration (capillary plasma referenced).
The GlucoMen® areo system is calibrated using
capillary plasma values determined with a
Yellow Springs 2300 analyser (YSI). The YSI
analyser is calibrated (as a secondary reference
measurement procedure) using a series of YSI
standards (primary calibrators) which are taken
from the NIST (National Institute of Standards
and Technology, USA).
Temperature
compensation:
Automatic compensation using a built-in thermo
sensor.
Batteries: Two 3.0 V lithium batteries (CR2032).
Battery life: A minimum of 1000 tests or about 1 year
(about 1000 runs for 1 transmission/day & 3
tests/day).
Memory: 730 test results with markers, dates and times.
When the memory is full, the new results will
replace the oldest ones.
Data
management:
Before/after meals, exercise, check markers
available.
Alarms: Up to three settable acoustic alarms.
Data transfer: Via dedicated data cable, NFC, or a Bluetooth
device.
Averages: Over 1, 7, 14, 30, 60, 90 days.
Automatic turn
off:
- After 50 seconds of inactivity before the test
(sensor strip inserted into the meter, drop icon
flashing).
- After 45 seconds of inactivity after the test and
Error messages Err2, Err3, Err4, HI and LO.
- After 5 seconds for Error messages Err1, Err5,
Temperature icon and Battery icon.
Dimensions: 85.5 mm (L) x 56 mm (W) x 18.2 mm (H)
Weight: 46 g (without batteries)
Meter Operating
Conditions:
- Temperature: 5 - 45 °C (41 - 113 °F).
-
Relative Humidity: 20 - 90% (no dew condensation).
Meter Storage
Conditions:
- Temperature: -20 – 50°C (-4 – 122 °F).
- Relative Humidity: 20 - 90%.
Test strip Storage
Conditions:
- Temperature: 4 - 30°C (39.2 - 86 °F) (both for
unopened vial and after opening).
- Relative Humidity: 20 - 90% (both for unopened
vial and after opening).
Control Solution
Storage
Temperature:
4 - 30 °C (39.2 - 86 °F)
Operating
environment:
The GlucoMen® areo meter complies with elec-
tromagnetic compatibility standards and has
been tested for electrostatic discharge and
electromagnetic fields according to EN 60601-1-
2:2007 +AC:2010.
Classification according to IEC/EN 60601-1: 2006
+AC:2010.
Electromagnetic compatibility: the equipment
complies with the EN 60601-1-2 EMC require-
ments.
EU directive/
classification:
98/79/EC / Annex II, List B
Device Standard: ISO 15197:2003