SECUTEST ST BASE(10) / PRO and SECULIFE ST BASE(25)
12 Gossen Metrawatt GmbH
Integrated Test Sequences
The test instrument is equipped with pre-
configured, integrated test sequences which
are selected via the switch positions at the
orange rotary switch level.
Note
Availability of the individual integrated
test sequences depends on the test
instrument type (SECUTEST ST... or
SECULIFE ST...), the selected fea-
tures (order features) and the enabled
extensions (activations).
Refer to your order / test instrument
and data sheet for details.
The integrated test sequences are freely
assignable, i.e. they can be individually
assigned to the rotary switch positions
(because there are more integrated test
sequences than rotary switch positions). But
the test instrument is preconfigured upon
delivery.
Which integrated test sequences are
assigned to the rotary switch positions on
your test instrument upon delivery depends
on several factors: test instrument type
(SECUTEST ST... or SECULIFE ST...),
selected features and enabled extensions.
Due to the great variety of possible combi-
nations, a listing of all delivery statuses
would go beyond the scope of this docu-
mentation.
In order to provide you with an impression,
the delivery status of a standard test instru-
ment is listed below as an example.
SECUTEST ST PRO, German version,
default settings:
Auto = automatic detection
Details concerning test sequences can be
found in the operating instructions (see sec-
tion 3 on page 8).
Integrated Test Sequences
Switch Positions at Orange Rotary Switch Level
Rotary
switch
position
Standard/
test
sequence
Meas.
type
Con-
nection
type
Protec-
tion
category
Voltage
spec.
A1 EN 50699 Auto Auto
Auto
A2 EN 50678 Auto Auto
Auto
A3
VDE 0701-
0702
Auto
Auto
Auto
A4 IEC 60974-4 Active Auto
Auto
U(0) DC
A5 IEC 62353 Passive Auto BF
Auto
A6 EN 50699 Passive
Test
socket
Auto
A7 EN 50678 Passive
Test
socket
Auto
A8
EN 50699-
VLTG
VLTG
EL1
adapter
PC I
A9 EN 50699 Active Auto Auto