XUB Series 29666 V4
Operating Manual
14
Cycle abort modes
If the cycle is aborted, the reason for the failure will be displayed.
If the operator aborts the cycle, the following screen will be displayed.
If the Level sensor is set to ON, and the liquid drops below the required level during
operation, the cleaning cycle will stop and the following screen will be displayed.
The Grant digital ultrasonic bath will not operate again until the tank is filled to the
correct level.
If the power to the ultrasonic generator should fall below a certain level during
operation, the cleaning cycle will stop and the following screen will be displayed.