EasyManua.ls Logo

Green G36P - Setting of Alarm Limits; Figure 5-2: Settings of Alarm Limits

Green G36P
48 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Configuration & Commissioning
26
5.2 Setting of Alarm Limits
The route for setting the alarm limits is Tools menu\Service (Password level 1:
9490)\Alarm setup.
+
/
--
+
/
--
+
/
--
+
/
--
+
/
--
Figure 5-2: Settings of alarm limits
The G36p has three alarms for oxygen
level:
Low Alarm level sets the low O
2
level
alarm – default set to 0.0 %.
High Alarm level sets the high O
2
level
alarm – default set to 22.0 %.
High-High Alarm lv sets the high O
2
level alarm – default set to 25.0 %.
Besides, there are other settings for
alarms as follows:
H&L Al. delay 1/10s sets the alarm de-
lay for both Low and High O
2
level
alarms. Default
set to 100 (equal to 10
seconds).
HH Al. delay 1/10s
sets the delay for
High-High O
2
alarm. Default set to 100
(equal to 10 seconds).
Analyzer alarms
is used to turn
ON/OFF the alarm function of the ana-
lyzer including analyzer
failure alarm
and oxygen level alarms.
When the Analyzer alarms is ON and
one of the alarm condition is present, the
Main Menu will display the alarm mes-
sage (e.g. Alarm: Low O
2
) in the bottom
left corner of the screen. Then, the alarm
relay will be triggered after the delay pe-
riod is over. For alarm relay output func-
tion and connection, please following the
instruction in section 3.4.3.

Table of Contents