Error Messages The GSI 61 displays error messages when the PRINT AUDIOGRAM pushbutton
is pressed when the following conditions exist:
If there is currently no saved data, this error message will be flashed:
No Test Data Stored
If test data exists, the presence of the Remote/Printer Option board
is verified; if it is not present, this error message will be flashed:
Not Available
If the Remote/Printer Option board is present, a check will be done
to see if a print job is already in progress. If so, it will be cancelled.
If the Remote is currently in use, this error message will be flashed:
Invalid Selection
If all of the above is satisfied, a check is done to see if the printer is
connected and ready to receive data. If not, this error message will
be flashed:
Check Printer
When all of the above is satisfied, an informational message is
displayed while outputting data to the printer:
Printing
If, during the course of printing, communications problems occur,
this error message will be flashed:
No Printer Response