Character Number of Field Name Field Description
Offset Characters
19
4
Ch 1 Numeric “-200” to “1000”
HL Display Data Range: -20.0 to 100.0 dB
23
1
Ch 1 Interrupt “0” = Off
“1” = On
24
4
Ch 2 Numeric “-200” to “1000”
HL Display Data Range: - 20. 0 to 100.0 dB
28
1
Ch 2 Interrupt “0” = Off
“1” = On
29
5
Frequency Display “__125” to “20000”
Data Range: 125 to 20000 Hz in
standard audiometric frequencies
34
5
Timer Value Elapsed time in seconds.
Range: “____0” to “12000”
seconds
39
3
Scorer # Correct Number of correct Speech Test
Value Zero responses.
Range: “__0” to “100”
42
3
Scorer Total Value Total number of Speech Test
presentations.
Range: “__0” to “100”
45
2
Checksum See page 6-4; Checksums
47
2
Record Terminator “CR,” “LF”
Contains all left and right ear test data currently stored in the test data memory. This
record is not GSI 16 compatible.
Character Number of Data Type Field Name
Offset Characters
Output Record
Type 6 - Test
Battery
Data Record
Recor
d
Prefix
Record Type
Pure Tone Test -
125 Hz Air
Conduction
Threshold- Test Ea
r
Field Description
“:”
“6”
-20 to 120 dB x 2
N
R = 260 to 540
N
T =32768 (0x8000)
0
1 ASCII
1
1
ASCII
Record Prefix
2
4
slnt
Left Ear Test Data –
Pure Tone
6
4
slnt
6-18
Pure Tone Test - -20 to 120 dB x 2
125 Hz Air NR = 260 to 540
Conduction NT = 32768 (0x8000)
Threshold-
Mas
k
ing Ea
r
1761-0100 Rev. B