Chapter 6 - Remote Options
Character Number of Data Type Field Name Field Description
Offset Characters
10
4
slnt Pure Tone Test - -20 to 120
125 Hz Bone NR = 260 to 540
Conduction NT = 32768 (0x8000)
Threshold- Test Ear
14
4
slnt Pure Tone Test - -20 to 120 dB x 2
125 Hz Bone NR = 260 to 540
Conduction NT = 32768 (0x8000)
Threshold-
Masking Ear
18
4
slnt Pure Tone Test - -20 to 120 dB x 2
125 Hz Sound Field NR = 260 to 540
NT = 32768 (0x8000)
22
20
Pure Tone Test - See Pure Tone Test -
250 Hz 125 Hz fields
42
20
Pure Tone Test - See Pure Tone Test -
500 Hz 125 Hz fields
62
20
Pure Tone Test - See Pure Tone Test -
750 Hz 125 Hz fields
82
20
Pure Tone Test - See Pure Tone Test -
1 kHz 125 Hz fields
102
20
Pure Tone Test - See Pure Tone Test -
1.5 kHz 125 Hz fields
122
20
Pure Tone Test - See Pure Tone Test -
2 kHz 125 Hz fields
142
20
Pure Tone Test - See Pure Tone Test -
3 kHz 125 Hz fields
162
20
Pure Tone Test - See Pure Tone Test -
4 kHz 125 Hz fields
182
20
Pure Tone Test - See Pure Tone Test -
6 kHz 125 Hz fields
202
20
Pure Tone Test - See Pure Tone Test -
8 kHz Low Freq 125 Hz fields
222
20
Pure Tone Test - See Pure Tone Test -
12 kHz 125 Hz fields
242
20
Pure Tone Test - See Pure Tone Test -
8 kHz High Freq 125 Hz fields
262
20
Pure Tone Test - See Pure Tone Test -
9 kHz 125 Hz fields
282
20
Pure Tone Test - See Pure Tone Test -
10 kHz 125 Hz fields
302
20
Pure Tone Test - See Pure Tone Test -
11.2 kHz 125 Hz fields
6-19