Section 3
3-13
3.7. MAIN MENU
▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒
Status Line
Last used configuration data
▒══════════╤══════════╤═════════╤══════════▒
Measure Setup VoltmeterSofcal
▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒
The main menu, on power up will display a null measurement configuration where all parameters
have been initialized to zero.
On power up the 6675A will default to a standby condition with the measurement cycle of
operations OFF i.e. <Measure off>.
During instrument operation, the operator can return to the main menu from any previously selected
lower menu level. When at the Main Menu level, pressing the <CANCEL> key will automatically
stop the measurement cycle and cause a return to the <Measure off> mode.
There are four other menu sequences that will also turn <Measure off> automatically:
1. When entering another measurement mode using the <Setup> <Edit>
<Mode> menu.
For example, when currently measuring a pair of 4 wire resistors and then
selecting 2 wire mode, the measurement cycle will be terminated and the
6675A will enter the <Measure off> mode.
2. When entering the <Voltmeter> menu from the main menu.
3. When changing the measurement type in the <Setup> menu from <Probe> to
<Resistor> or vice versa.
4. When entering the <Sofcal> <Voltmeter> menu or the <Sofcal>
<Diagnostic> menu.