Section 3
3-14
3.7.1.1. <Measure> MENU Fn-1 (MAIN) Page 1 of 3.
▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒
▒STATUS LINE ▒
▒ ■▒
▒Current reading or ▒
▒Measurement configuration ▒
▒ ▒
▒ ▒
▒──────────┬──────────┬─────────┬──────────▒
▒Ratio │Meas Off │Update 1 │MORE 1-3 ▒
▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒
Ohms Meas On Update 2
Celsius Update 4
Fahrenheit
kelvin
Ratio
Ohms
The <STATUS LINE> will only appear if the 6675A is currently measuring a set of
resistors. The information displayed in the <STATUS LINE> depends on the
options selected.
1. if <PPM>, <PPM SP>, <DELTA>, or <DELTA SP> is the current selection
then the <STATUS LINE> will contain the Datum of proceding numbers.
2. if <N AVERAGE> is selected the the standard deviation will be displayed,
on the <STATUS LINE>, for the raw unfiltered values.
3. if the current measurement mode is <I EXTEND> then the extender ratio
will also be displayed on the <STATUS LINE>.
Available functions (menu page 1 of 3).
Function key <Fn-1> toggles through the set of available measurement units for the
displayed result; when the <Resistor> measurement mode is active, the displayed
result can have units of <Ratio> and <Ohms>. When the <Probe > measurement
mode is active, the displayed result can have additional units of temperature,
<Celsius>, <Fahrenheit> and <kelvin>. The active engineering units will be
displayed in the <Fn-1> function key descriptor position.
Function key <Fn-2> enables or disables the measurement cycle by turning
<Measure ON> or <Measure OFF>. The state of the measurement cycle is displayed
in the <Fn-2> function key window.