Section 3
3-15
NOTE:
A graphic block ( ) will be displayed on the right most vertical column of the
menu to show which measurement cycle the 6675A is in. This block will appear
on all menus and is one way to tell if the 6675A is making a measurement.
Function key <Fn-3> selects the screen measurement update rate as a function of the
test current reversal rate. <Update 1> sets a screen update rate at twice the test
current reversal rate; <Update 2> sets a screen update rate at the test current reversal
rate; <Update 4> sets a screen update rate at half the test current reversal rate.
Function key <Fn-4> selects the next menu page of available functions.