Section 3
3-16
3.7.1.2. <Measure> MENU Fn-1 (MAIN) Page 2 of 3.
▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒
STATUS LINE
CURRENT READING OR
Measurement configuration
▒──────────┬──────────┬─────────┬──────────▒
Power x1 No Filter Normal MORE 2-3
▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒
Power x2 Decimation PPM
N Average Delta
PPM SP
Delta SP
Available functions (menu page 2 of 3).
Function key <Fn-1> selects twice the rated power.
Function key <Fn-2> turns on the filter function to operate on the measurement data.
Select from <No Filter>, <Decimation> and <N Average> where the sample size N
must be entered from the keypad.
Function key <Fn-3> selects the mode of displaying the measurement data as
<Normal>; parts per million <PPM>; change <Delta>; change ppm from a setpoint
<PPM SP>; and change from a setpoint <Delta SP>. When this function is enabled,
the datum value will be displayed on the STATUS LINE of the display.
Function key <Fn-4> displays next menu page of available functions.