Section 3
3-21
Function Key <Fn-3> ,<Update>, allows selection of the number of times that the
display should be updated for each measurement cycle; 1, 2, or 4 times per
measurement cycle.
<Edit> Resistors MENU Fn-2 (Resistors_Setup_MAIN)
Edit resistors menu page 2 of 2.
▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒
▒Mode: RS
: ▒
▒Display: RS SN: ▒
▒Update: RX
: ▒
▒Filter: <Fn-1> RRate: ▒
▒Power: <Fn-2> Ratio|mAmps|Volts: ▒
▒ ▒
▒──────────┬──────────┬─────────┬──────────▒
▒Filter │Power │ │MORE 2-2 ▒
▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒
No Filter Power x1
Decimation Power x2
N Average
Function Key <Fn-1> ,<Filter>, allows selection of the filter that will be used on the
displayed readings, where:
No Filter = display most recent reading
Decimation = display (100-N) percent of the previous
reading. Where N is the percentage value set
in the <Sofcal><Filter> menu for Decimation.
N Average = display the average of the last N readings.
Where N is the is the number of readings to be
averaged and is that value set in the
<Sofcal><Filter> menu.
Function Key <Fn-2> ,<Power>, allows selection of a multiple of the power to be
dissipated in the test resistor as X1 (unity) or as X2 (two times power level).