3.7.2.2.1.1. <Mode> MENU Fn-1 (Edit_Resistors_Setup_MAIN)
▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒
▒Mode: <Fn-??> RS : ▒
▒Display: RS SN: ▒
▒Update: RX : ▒
▒Filter: RRate: ▒
▒Power: Ratio: ▒
▒ ▒
▒══════════╤══════════╤═════════┬──────────▒
▒4 Terminal│2 Terminal│I Extend │ ▒
▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒
Function key <Fn-1>, <4 Terminal>, will select the 4-terminal measurement
mode of operation.
Function key <Fn-2>, <2 Terminal>, will select the 2-terminal measurement
mode of operation.
Function key <Fn-3>, <I Extend>, will select operation using a high current
range extender for low value resistance measurements.
3.7.2.2.1.1.1. <4 Terminal> MENU Fn-1
(Mode_Edit_Resistors_Setup_MAIN)
Enter configuration for 4-terminal resistance mode measurements.
▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒
▒Resistance RS: ▒
▒Serial No. RS: ▒
▒Resistance RX: ▒
▒Reversal Rate: ▒
▒Test Current (mA): ▒
▒Max Current (mA): ▒
▒──────────┬──────────┬─────────┬──────────▒
▒Edit │Ok │ │ ▒
▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒
Press <Edit> and use arrow keys to move the highlight to the data line to be changed.
Modify data and press <Ok>. Check resistance limits for Rs and Rx. Check also
power limits set by test current.
Where the Test Current (I(test) Ix) sets the current in the UNKNOWN
RESISTOR (Rx).